Agilent Technologies N8900 series Operating Manual page 58

Autoranging system dc power supply
Table of Contents

Advertisement

Front Panel Menu
Select Transient\Control.
Scroll to Initiate. Then press Select.
It takes a few milliseconds for the instrument to be ready to receive a trigger signal after receiving the INITi-
ate:TRANsient command. If a trigger occurs before the trigger system is ready for it, the trigger will be ignored. You can
test the WTG_tran bit in the operation status register to know when the instrument is ready to receive a trigger.
Front Panel Menu
Select Transient\Control.
The Trig state field indicates "Initiated".
If a bit value of 16 is returned in the query, the WTG_tran bit is true, and the instrument is ready to receive the trigger
signal. Refer to
Status Tutorial
Trigger the output step
The trigger system is waiting for a trigger signal in the initiated state. Trigger the transient as follows:
Front Panel Menu
Select Transient\Control.
Select Trigger to generate an immediate
trigger signal regardless of the trigger
source setting.
If the trigger does not occur, you must manually return the trigger system to the idle state. The following commands
return the trigger system to the idle state:
Front Panel Menu
Select Transient\Control.
Then select the Abort control.
When a trigger is received, the triggered functions are set to their programmed trigger levels. When the triggered
actions are completed, the trigger system returns to the idle state.
You can test the TRAN_active bit in the Operation Status register to know when the transient trigger system has
returned to the idle state.
Front Panel Menu
Select Transient\Control.
The Trig state field indicates "Idle".
If a bit value of 64 is returned in the query, the TRAN_active bit is true, and the transient action is NOT complete. When
the TRAN_active bit is false, the transient action is complete. Refer to
Agilent N8900 Series Operating and Service Guide
To initiate the transient trigger system:
INIT:TRAN
To query the WTG_tran bit (bit 4):
STAT:OPER:COND?
for more information.
To generate a transient trigger:
TRIG:TRAN
Alternatively, you can also program a
*TRG or an IEEE-488 <get> command.
ABOR:TRAN
To query the TRAN_active bit (bit 6):
STAT:OPER:COND?
SCPI Command
SCPI Command
SCPI Command
SCPI Command
SCPI Command
Status Tutorial
Programming Output Steps
for more information.
57