Command Quick Reference

Command Quick Reference
Some [optional] commands have been included for clarity. All settings commands have a corresponding query. See the
Syntax Conventions
for SCPI.
ABORT Command
ABORt
      :TRANsient
CALibrate Commands
CALibrate
      :COUNt?
      :CURRent
[:LEVel] <value>
      :DATA <value>
      :DATE <"date">
      :LEVel P1|P2|P3|P4
      :PASSword <value>
      :SAVE
      :STATe 0|OFF|1|ON
      :VOLTage
[:LOCal]
[:LEVel] <value>
            :REMote <value>
[:LEVel] <value>
DISPlay Command
DISPlay
[:WINDow]
[:STATe] 0|OFF|1|ON
IEEE-488 Commands
*CLS
*ESE <value>
*ESR?
*IDN?
*LRN?
*OPC
*OPT?
104
Cancels any triggered actions.
Returns the number of times the unit has been calibrated.
Calibrates the current programming and measurement.
Enters the calibration value read by the external meter.
Enters the calibration date in nonvolatile memory.
Advances to the next level in the calibration.
Sets a numeric password to prevent unauthorized calibration.
Saves the calibration constants in non-volatile memory.
Enables or disables calibration mode.
Calibrates the local voltage programming and measurement.
Calibrates the remote voltage programming and measurement.
Turns the front panel display on or off.
Clear status
Standard event status enable
Return event status register
Return instrument identification
Returns a sequence of SCPI commands
Enable "operation complete" bit in ESR
Return option number
Agilent N8900 Series Operating and Service Guide