Gpib Cable Replacement; Tools Required; Removal Procedure; Replacement Procedure - Agilent Technologies E4980A Service Manual

Precision lcr meter
Hide thumbs Also See for E4980A:
Table of Contents

Advertisement

GPIB Cable Replacement

Tools Required

Torque screwdriver, TORX T20
Box torque wrench, 7 mm

Removal Procedure

Refer to Figure 6-13 for this procedure.
Step 1. Remove the 3/4 Rack Module Top as described in "3/4 Rack Module Top Replacement" on
page 90.
Step 2. Disconnect the GPIB Cable connector (item 1) from PPM PCA card.
Step 3. Remove the two 7mm standoff (item 2) fastening the GPIB cable.
Step 4. Remove the GPIB Cable (item 3) from the Rear Panel Assembly.

Replacement Procedure

Step 1. Reverse the order of the removal procedure.
NOTE
When you fasten the GPIB Cable with the two 7 mm standoff, set box torque wrench's
torque to 8.7 in-lb.
Chapter 6
Replacement Procedure
GPIB Cable Replacement
113