TEMP?
Description
Remarks
Example
Temperature Query. Returns the multimeter's internal temperature in
degrees Centigrade.
Syntax
TEMP?
•
Monitoring the multimeter's temperature is helpful to determine when to
perform autocalibration.
•
Related Commands: ACAL, CAL, CALSTR
10 OUTPUT 722; "TEMP?"
20 ENTER 722; A
30 PRINT A
40 END
!READ TEMPERATURE
!ENTER RESULT
!PRINT RESULT
Chapter 5 Command Summary