SCAL

Description
Remarks
Example
76 Chapter 5 Command Summary
Service Calibration. Adjusts the AC sections of the instrument. Calculates
the corrections to accurately measure frequency and adjusts the ac ranges.
The SCAL command is located in the full command menu.
Note
The SCAL command is used in the AC adjustment procedure of Chapter 3
and the procedure must be performed in the order specified.
Syntax
SCAL value [,security_code]
Specifies the value of the adjustment source that will be used to do the
value
service adjustment of the multimeter. Valid choices for value are 1E5, 10,1,
and 0.1. 1E5 performs a frequency calibration while 10, 1, and 0.1 do ac
range adjustment.
security_code
SECURE command, you must enter the correct security code to perform an
SCAL. If SCAL is not secured (security code = 0), no security code is
required to execute SCAL. Refer to the SECURE command for more
information on the security code and how to secure the adjustment of the
3458A.
The SCAL command is in the full menu; it is not in the short menu.
With a 100 KHz input, SCAL 1E5 calculates the constants allowing the
multimeter to indicate 100 KHz.
SCAL 10, SCAL 1, and SCAL 0.1 do ac range calibration calculating
high frequency dac and interpolator constants.
For fastest calibration with SCAL, the multimeter can be left in the DCV
function.
Related Commands: ACAL, CAL, SECURE
OUTPUT 722;"SCAL 1E5"
OUTPUT 722;"SCAL 10"
OUTPUT 722;"SCAL 1"
OUTPUT 722;"SCAL 0.1"
When a security code is set to a number other than 0 by the
!Adjusts for frequency
!Adjusts range