This Appendix contains technical information about all the glove boxes including
electrical specifications and environmental operating conditions.
All models: 12 Amps, 100-115V or 230V, 50/60 Hz
Vacuum pump exterior receptacle rated at 12 Amps, 100-115V, 50/60 Hz or 6
Amps, 230V 50/60 Hz.
Single width units require two line cords and double width units require three
line cords.
Environmental Conditions
 Indoor use only.
 Maximum altitude: 6562 feet (2000 meters).
 Ambient temperature range: 41° to 120°F (5° to 49°C).
 Maximum relative humidity: 80% for temperatures up to 88°F (31°C),
decreasing linearly to 50% relative humidity at 104°F (40°C).
 Main supply voltage fluctuations not to exceed ±10% of the nominal
voltage.
 Transient over-voltages according to Installation Categories II (Over-
voltage Categories per IEC 1010). Temporary voltage spikes on the AC
input line that may be as high as 1500V for 115V models and 2500V for
230V models are allowed.
 Used in an environment of Pollution degrees 2 (i.e., where normally only
non-conductive atmospheres are present). Occasionally, however, a
temporary conductivity caused by condensation must be expected, in
accordance with IEC 664.
Product Service 1-800-522-7658
C
:
C
:
t
i
o
n
s
t
i
o
n
s
63

Advertisement

loading
Need help?

Need help?

Do you have a question about the Protector 50600 and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

Protector 50601Protector 50800Protector 50801