Appendix D: Controlled Atmosphere Oxygen & Moisture Removal Rates
Protector CA (Fiberglass) Moisture Removal with AtmosPure Re-Gen Gas Purifier
1000
600
100
50.1
10.6
10
6.7
4.6
3.5
2.3
1
1
10
100
1000
10000
Time (Minutes)
This chart shows moisture level decrease from 600 ppm to 2.3 ppm after
*
9875 minutes. Two separate calibrated moisture monitors confirmed results.
*
Results were attained at Labconco Corporation under controlled conditions. Actual results
may vary.