Liquid level sight
gauge. Fill oil to
6" (152 mm) from
bottom.
Bottom of bubbler
14
Figure 3-1
Pressure Relief Bubbler Set-up
Consult with Safety
Officer for exhausting
3/4" OD to the outside.
Fill with 0.87 specific
gravity oil using a 3/4"
I.D. hose and funnel.