Chapter 5: Using Your Glove Box
Setup Menu – Operation Mode 4
Pressing the key opposite the word "Setup," allows access to the setup operation mode. In
the setup mode, the user may select their preferred units of pressure that are displayed,
restore settings back to the default state, or calibrate the pressure sensors. The keys
adjacent to the up and down arrows on the lower right of the display are used to select one
of the three options. After toggling up or down, push the NEXT key to select the option.
Back to
Previous
Screen
Option 1
Units of Measure: If this option is selected, the key adjacent to "Next" is used to
select pressure units of inches of water, inches of mercury or mBar.
Helpful conversions are as follows:
1 mBar = .39 inches of water column
1 mbar = .0292 inches of mercury
1 inch of water = 2.52 mBar = .0737 inches of mercury
1 inch of mercury = 34.25 mBar
All pressures are measured with respect to atmosphere pressure.