Pressure Change Method Leak Testing Summary
(ISO-10648-2)
The method consists of measuring the pressure rise per unit time after establishing
a negative pressure in the main chamber of the glove box. Use
-1000 Pa for the acceptance test and -250 Pa for the operational test. The hourly
leak rate includes adjustments for barometric pressure and temperature changes
which can be calculated as T
T
= (60/t) * ((PnT
/P
f
1
1
t is the duration of the test in minutes.
P
is the absolute pressure (ambient barometric pressure minus under pressure) at
1
the first reading in Pascals.
Pn is the absolute pressure (ambient barometric pressure minus under pressure) at
the last reading in Pascals.
T
is the temperature at the first reading in Kelvin.
1
Tn is the temperature at the last reading in Kelvin.
60 represents 60 minutes in an hour.
T in Kelvin is equal to degrees Celsius + 273.