Labconco Protector 50600 User Manual page 51

Controlled atmosphere glove boxes

Advertisement

Front panel
screw location
Window Replacement
 It is recommended to decontaminate before opening the glove box and to
check with your safety officer.
 Remove 30 screws on outer cover plate by using minor turns and across
from each other diagonally to reduce pressure evenly. Be careful as the
screws have stainless steel lube that will stain clothing. Hold lower frame
in and see if top frame will come off without glass. Frame plate may come
off with window, but do not depend on gasket material to hold glass. To
prevent injury to your feet or toes, do not let glass slide to floor. Glass sits
on a 1/4 inch ledge and should stay as long as it is not pulled out, but do not
depend on it. Clean inside of window after removal.
 If you change gasket(s), clean old glue off frame. Apply new glue to frame
and gasket (check old gasket when taking off so you will know which side
is up). Allow glue to be tacky before putting together. Start gasket at
Product Service 1-800-522-7658
Chapter 6: Maintaining Your Glove Box
Fluorescent
Bulbs
Fluorescent Bulb
Light Socket
Figure 6-1
Fluorescent Lights
Fluorescent Bulb
Spring Clip
47