Download Print this page

Labconco Protector 50600 User Manual page 48

Controlled atmosphere glove boxes

Advertisement

Chapter 5: Using Your Glove Box
Auto Pressure Controller Specifications
Glove Box
Pressure Control
Pressure Control Limits
Manual Control Available
Audio Alarm level
Glove Box Purge/Fill
Purge/Fill Pressure Limit
Manual Control Available
Number of Purge/Fill
Cycles
Audio Alarm level
Transfer Chamber
Purge/Fill
Purge Pressure Limit
Fill Pressure Limit
Number of Purge/Fill
Cycles
Audio Alarm level
44
Operating Range
-5 inches of water column to
+5 inches of water column;
[+/- 12.6 mBar]
Purge (Gas out);
Fill (Gas in)
Whenever pressure
outside limits
+/- 5 inches of water column;
[+/- 12.6 mBar].
Purge (Gas out);
Fill (Gas in)
0 to 499 – main chamber
Whenever pressure
outside limits
-29.5 inches of Mercury/Hg
[-999 mBars]
0 inches of Mercury/Hg
[0 mBars](atmosphere)
0 to 199 – transfer chamber
-29.5 inches of Mercury/Hg
[-999 mBars]
Product Service 1-800-522-7658
Default Settings
+/- 5 inches of water column;
[+/- 12.6 mBar].
+/- 5 inches of water column;
[+/- 12.6 mBar].
1
-29.5 inches of Mercury/Hg
[-999 mBars]
0 inches of Mercury/Hg
[0 mBars](atmosphere)
1
Whenever pressure/vacuum
outside limits