14.12.3 Models BCA5201 | BCA2201
Scale interval (d)
Maximum capacity (Max)
Repeatability
At 5%, typical value
At approx. the maximum load, typical value
Linearity deviation
Limits
Typical value
Sensitivity drift between +10°C and +30°C
Tare maximum capacity (subtractive)
isoCAL (only for I-1x models):
Temperature change
Time interval
Only for models with approval:
Accuracy class
Type
Verification scale interval (e)
Minimum load (Min)
Minimum initial weight according to USP (United
States Pharmacopeia), Chap. 41
Optimum minimum initial weight
Typical minimum initial weight
Typical measurement time
Typical stabilization time
Recommended calibration weight
External calibrated test weight
Accuracy class in accordance with OIML R111-1
Unit
Value
Value
mg
100
100
g
5,200
5,200
mg
50
50
mg
100
50
± mg
300
100
± mg
100
60
±
7
2
ppm/K
<100% of maximum capacity
K
2
h
6
II
BC-AE
mg
100
mg
5,000
g
82
82
g
100
100
s
≤1.0
≤1.0
s
≤0.9
≤0.9
g
5,000
5,000
F2
F2
Entris
II Advanced Line Operating Instructions
®
Technical Data
Value
Value
100
100
2,200
2,200
50
50
100
50
300
100
100
60
7
2
2
6
II
BC-AE
100
5,000
82
82
100
100
≤1.0
≤1.0
≤0.9
≤0.9
2,000
2,000
F2
F2
53