System Settings
7.1
Performing System Settings
Default settings can be adjusted for the device and the
applications in order to align with the ambient
conditions and individual operating requirements. You
can change the settings at any time.
The following settings are necessary to operate the
device together with connected components:
− Set up the communication of the connected devices
− Set up additional components
When using the device for the first time and after
resetting the device settings, the installation assistant
starts. This assists with the following settings:
− Set the menu language
− Set the date and time
Procedure
Press the [Main menu] button.
t
To adjust settings: Open the desired menu.
t
Select and confirm the desired parameter
t
(parameters, see Chapter "4.10 "Main Menu" Menu
Structure", page 17).
Exit the menu.
t
7.2
Setting the Calibration and
Adjustment
7.2.1
Switching the isoCAL Function On or
Off (Only Model I- x)
When using the isoCAL function, the device performs
an automatic time- and temperature-dependent
internal calibration and adjustment.
Depending on the device model, other temperature
ranges may apply for ensuring metrological data when
the isoCAL function is switched off (see Chapter "14.1
Ambient Conditions", page 47).
If this relates to a conformity-assessed device
in legal metrology: In some cases it is not
possible to switch off the isoCAL function.
Procedure
Open the main menu.
t
Open the "Settings" / "Cal.-/adj. settings" / "isoCAL"
t
menu.
To set the automatic start of the isoCAL function:
t
Select "Automatic" setting.
To set the manual start of the isoCAL function: Select
t
the "Info, manual start" setting.
To switch off the isoCAL function: Select the "OFF"
t
setting.
7.3
Editing User Management
7.3.1
Setting and Changing Admin Password
Requirements
The "Administrator" is logged into the device.
Procedure
Open the "Settings" / "User management" / "Set
t
admin password" menu.
Press the [...] button.
t
The input screen for the admin password is displayed.
y
Enter the desired password in the input field and
t
confirm with the [Confirm] button.
The input screen for the user password is displayed
y
again.
Enter the password in the input field a second time
t
and confirm with the [Confirm] button.
Press the [Confirm] button.
t