14.12 Metrological Data
14.12.1 Models BCA6202 | BCA4202 | BCA3202
Scale interval (d)
Maximum capacity (Max)
Repeatability
At 5%, typical value
At approx. the maximum load, typical value
Linearity deviation
Limits
Typical value
Sensitivity drift between +10°C and +30°C
Tare maximum capacity (subtractive)
isoCAL (only for I-1x models):
Temperature change
Time interval
Only for models with approval:
Accuracy class
Type
Verification scale interval (e)
Minimum load (Min)
Minimum initial weight according to USP (United
States Pharmacopeia), Chap. 41
Optimum minimum initial weight
Typical minimum initial weight
Typical measurement time
Typical stabilization time
Recommended calibration weight
External calibrated test weight
Accuracy class in accordance with OIML R111-1
Unit
Value
mg
10
g
6,200
mg
5
mg
10
± mg
20
± mg
6
±
2
ppm/K
<100% of maximum capacity
K
2
h
6
II
BC-AE
mg
100
mg
500
g
8.2
g
10
s
≤1.0
s
≤0.9
g
5,000
F1
Entris
II Advanced Line Operating Instructions
®
Technical Data
Value
Value
10
10
4,200
3,200
5
5
10
10
20
20
6
6
2
2
2
2
6
6
II
II
BC-AE
BC-AE
100
100
500
500
8.2
8.2
10
10
≤1.0
≤1.0
≤0.9
≤0.9
2,000
2,000
F1
F1
51
Need help?
Do you have a question about the Entris II Advanced Series and is the answer not in the manual?