21.3.2 Sending an operating system to a remote data
logger
CAUTION:
Sending an OS remotely to a data logger with Operating System 4 or earlier is not
recommended. These data loggers should be updated using the previous instruction
(Sending an operating system to a local data
If you have a data logger that is already deployed, you can update the OS over a
telecommunications link by sending the OS to the data logger as a program. In most instances,
sending an OS as a program preserves settings. This allows for sending supported operating
systems remotely (check the release notes). However, this should be done with great caution as
updating the OS may reset the data logger settings, even settings critical to supporting the
telecommunication link.
The default.CR300 program can be edited to preserve critical data logger settings such as
communications settings. See
 1. Download the latest Operating System at
 2. Locate the .exe download and double-click to run the file. This will extract the .obj OS file to
the C:\Campbellsci\Lib\OperatingSystems folder.
 3. Using data logger support software, connect to your data logger.
 
LoggerNet users, select Main and click Connect
l
the data logger from the Stations list, then click Connect
 
PC400 users, select the data logger from the list and click Connect
l
 4. Select File Control
 5. Click Send
at the top of the File Control window.
 6. Navigate to the C:\Campbellsci\Lib\OperatingSystems folder.
Default
program (p. 143) for more information.
at the top of the Connect window.
.
logger (p. 164)).
https://www.campbellsci.com.au/downloads
on the LoggerNet toolbar, select
.
.
21. CR300 series maintenance     165
.