CRBasic Example 1: Four-wire full-bridge measurement and processing
Public
X_1
Public
R_1
Public
R_2 = 1000
Public
R_3 = 1000
Public
R_4 = 1000
'Main Program
BeginProg
Scan(500,mSec,1,0)
'Full Bridge Measurement:
BrFull(X,1,mV2500,1,Vx1,1,2500,False,True,0,60,1.0,0.0)
X_1 = ((-1 * X) / 1000) + (R_3 / (R_3 + R_4))
R_1 = (R_2 * (1 - X_1)) / X_1
NextScan
EndProg

18.3.2 RTD and PRT

RTDs (resistance temperature detectors) are resistive devices made of platinum, nickel, copper,
or other material. Platinum RTDs, known as PRTs (platinum resistance thermometers) are very
accurate temperature measurement sensors.
A PRT element is a specialized resistor with two connection points. Most PRTs are either 100 Ω or
1000 Ω. This number is the resistance the PRT has at 0 °C. The resistance of a PRT increases as it is
warmed. Industry standards define how PRTs respond to temperature.
BrHalf4W()
in combination with
measuring RTDs.
See the CRBasic Editor help for detailed instruction information and program examples:
https://help.campbellsci.com/crbasic/cr300/

18.3.3 Strain measurements

A principal use of the four-wire full bridge is the measurement of strain gages in structural stress
StrainCalc()
analysis.
configuration used. All strain gages supported by
'Quarter-bridge', 'half-bridge' and 'full-bridge' refer to the number of active elements in the
bridge schematic. In other words, a quarter-bridge strain gage has one active element, a half-
bridge has two, and a full-bridge has four.
StrainCalc()
requires a bridge-configuration code. The following table shows the equation
used by each configuration code. Each code can be preceded by a dash (-). Use a code without
'Resistance of fixed resistor R2
'Resistance of fixed resistor R3
'Resistance of fixed resistor R4
PRTCalc()
calculates microstrain (µɛ) from the formula for the specific bridge
are the recommended CRBasic instructions for
.
StrainCalc()
use the full-bridge schematic.