Operation options
36
6
Operation options
6.1
Human interface and display elements for FEI51,
FEI52, FEI54, FEI55
8
1
22
FEI51, FEI52, FEI54, FEI55 human interface
1
Key
2
Green LED - operational status
3
Green LED
4
Red LED - fault
5
Green LED
6
Green LED
7
Yellow LED - switching state
8
Mode switch
9
Key
1. Operation - select for normal operation
2. Restor factory settings:
press and for 20 s restore factory settings
3. Calibration
press to set empty calibration
press to set full calibration
press and for 10 s to reset the calibration and switch-point adjustment
4. Switch-point adjustment
press to decrease the switch-point
press to increase the switch-point
5. Measuring modes
press to decrease the measuring range
press once to set the two-point control Δs
press twice to activate the build-up mode
6. Switching delay
press to decrease the delay
press to increase to delay
7. Self-test
press and to activate the self-test
8. Setting MIN/MAX failsafe mode or SIL mode
press for minimum
press for maximum
press and to lock or unlock the SIL mode
2
3 4 5 6
7
7
3
-
+
FEI5x
5
-
+
1
2
Liquicap M FTI51
9
A0042394
Endress+Hauser