Technical data
80
12.4
Performance characteristics
According to DIN 61298-2
• Uncertainty: maximum ±0.3 %
• Non-repeatability: maximum ±0.1 %
12.4.1
Ambient temperature effect
Electronic insert
< 0.06 % per 10 K related to the full-scale value
Separate housing
capacitance change of connecting cable per meter 0.15 pF per 10 K
12.5
Operating conditions: Environment
12.5.1
Ambient temperature range
• F16 housing: –40 to +70 °C (–40 to +158 °F)
• remaining housing: –50 to +70 °C (–58 to +158 °F)
• observe derating
• use a protective cover, when operating outdoors
12.5.2
Climate class
DIN EN 60068-2-38/IEC 68-2-38: Z/AD check
12.5.3
Vibration resistance
DIN EN 60068-2-64/IEC 68-2-64: 20 to 2 000 Hz, 0.01 g
12.5.4
Shock resistance
DIN EN 60068-2-27/IEC 68-2-27: 30 g acceleration
12.5.5
Cleaning
Housing:
Make sure that the housing surface and seals are resistant to cleaning agents.
Probe:
Depending on the application, buildup of contamination or soiling can form on the probe.
A high level of material buildup can affect the measurement result.
The regular cleaning of the probe is recommended if the medium tends to create a high
level of buildup.
Make sure that the insulation of the probe is not damaged if hosing down or during
mechanical cleaning.
12.5.6
Degree of protection
All protection degree regarding EN60529.
Type4X protection degree regarding NEMA250.
Liquicap M FTI51
2
/Hz
Endress+Hauser