Technical data
84
Probe ⌀22 mm (0.87 in) including insulation
–1 to 50 bar (–14.5 to 725 psi)
Refer to the following standards for the pressure values permitted at higher temperatures:
• EN 1092-1: 2005 Table, Appendix G2
With regard to its resistance and temperature property, the material 1.4435 is identical
to 1.4404 (AISI 316L) which is grouped under 13E0 in EN 1092-1 Tab. 18. The
chemical composition of the two materials can be identical.
• ASME B 16.5a - 1998 Tab. 2-2.2 F316
• ASME B 16.5a - 1998 Tab. 2.3.8 N10276
• JIS B 2220
The lowest value from the derating curves of the device and the selected flange applies.
12.6.3

Pressure and temperature derating

For process connections ½", ¾", 1", flanges <DN50, <ANSI 2", <JIS 10K
(10 mm (0.39 in) and 14 mm (0.55 in) rod) as well as process connections ¾", 1",
flanges <DN50, <ANSI 2", <JIS 10K (16 mm (0.63 in) rod)
Rod insulation: PTFE, PFA
[psi]
[bar]
P
p
+362.5
+25
+29
+2
–14.5
–1
–80
–112
 64
Process presurre and temperature derating diagram
P
Process pressure
p
T
Process temperature
p
For process connections 1½", flanges ≥DN50, ≥ANSI 2", ≥JIS 10K (⌀ 16 mm (0.63 in)
rod)
Rod insulation: PTFE, PFA
+50
+122
Liquicap M FTI51
[ ]
°C
+200
T
P
[°F]
+392
A0043640
Endress+Hauser