Aligning The Housing; Sealing The Probe Housing; Post-Installation Check - Endress+Hauser Liquicap M FTI51 Operating Instructions Manual

Capacitance point level switch for liquids
Hide thumbs Also See for Liquicap M FTI51:
Table of Contents

Advertisement

Liquicap M FTI51
Endress+Hauser
 16
Spring washer mounting
1
Spring washer
4.5.2

Aligning the housing

The housing can be rotated 270 ° to align the cable entry. To prevent moisture penetration,
route the connecting cable downwards in front of the cable gland and secure it with a cable
tie. This is particularly recommended for outdoor mounting.
Aligning the housing
1 →
PH-2
A0042107
Loosen the clamping screw.
The clamping screw for aligning the housing type T13 is located in the electronics
compartment.
4.5.3

Sealing the probe housing

Make sure that the cover is sealed. Water cannot enter into the device when performing
installation, connection and configuration tasks. Always seal the housing cover and cable
entries securely.
The O-ring seal on the housing cover is shipped with a coat of special lubricant applied. In
this way, the cover can be sealed tight and the aluminum thread does not bite when
screwing down.
Never use mineral oil-based grease as this destroys the O-ring.
4.6

Post-installation check

After installing the measuring device, carry out the following checks:
 Do a visual check for damages.
1
2 →
...270°
A0042108
Align the housing into the
needed position.
Mounting
A0040477
3
PH-2
A0042109
Tighten the clamping screw
with torque
< 1 Nm (0.74 lbf ft).