Liquicap M FTI51
Endress+Hauser
16
Spring washer mounting
1
Spring washer
4.5.2
Aligning the housing
The housing can be rotated 270 ° to align the cable entry. To prevent moisture penetration,
route the connecting cable downwards in front of the cable gland and secure it with a cable
tie. This is particularly recommended for outdoor mounting.
Aligning the housing
1 →
PH-2
A0042107
‣
Loosen the clamping screw.
The clamping screw for aligning the housing type T13 is located in the electronics
compartment.
4.5.3
Sealing the probe housing
Make sure that the cover is sealed. Water cannot enter into the device when performing
installation, connection and configuration tasks. Always seal the housing cover and cable
entries securely.
The O-ring seal on the housing cover is shipped with a coat of special lubricant applied. In
this way, the cover can be sealed tight and the aluminum thread does not bite when
screwing down.
Never use mineral oil-based grease as this destroys the O-ring.
4.6
Post-installation check
After installing the measuring device, carry out the following checks:
Do a visual check for damages.
1
2 →
...270°
A0042108
‣
Align the housing into the
needed position.
Mounting
A0040477
3
PH-2
A0042109
‣
Tighten the clamping screw
with torque
< 1 Nm (0.74 lbf ft).
23