Endress+Hauser Liquicap M FTI51 Operating Instructions Manual page 20

Capacitance point level switch for liquids
Hide thumbs Also See for Liquicap M FTI51:
Table of Contents

Advertisement

Mounting
20
7 →
A0042117
Remove the blade plug from
the socket.
 13
Cable connections
1
External screening (not required)
2
Strand black (bk) (not required)
3
Coaxial cable with central core and screen
4
Solder the red (rd) strand with the central core of the coaxial cable (probe)
5
Insulated strand (ye) with the heat shrinking sleeve
6
Strand yellow and green (gn/ye) with a ring terminal
Probe with active buildup compensation
Disconnecting the connection cable
1 →
22
A0042111
Loosen the pressing screw with
an open-end wrench AF22.
8 →
A0042546
Loosen the screw to disconnect
the yellow-green cable.
1
bk
2
3
6
4
5
gn/ye
rd
ye
2 →
A0042112
Pull the insert seal out of the
cable gland.
Liquicap M FTI51
9
A0042119
Loosen the nut (M4) of the
blade plug.
A0042544
3 →
22
34
A0042113
Block the adapter disk with the
open-end wrench AF34 and
loosen the cable gland with the
open-end wrench AF22.