Repair

10.1

General notes

The Endress+Hauser repair and conversion concept provides the following:
• The measuring devices have a modular design
• Spare parts are grouped into logical kits with the associated Installation Instructions
• Repairs are carried out by Endress+Hauser Service or by appropriately trained customers
• Certified devices can only be converted to other certified devices by Endress+Hauser
Service or at the factory
10.2

Spare parts

Find spare parts
Check whether it is possible to use the spare part for the measuring device.
1. Launch the Endress+Hauser Device Viewer via a web browser:
www.endress.com/deviceviewer
2. Enter the order code or the product root in the respective field.
 Once the order code or the product root has been entered, all the suitable spare
parts are listed.
The product status is displayed.
Available drawings of the spare parts are displayed.
3. Locate the order code of the spare part set (on the product label on the package).
 NOTE!
The order code of the spare part set (on the product label on the package) can
differ from the production number (on the label directly on the spare part)!
4. Check whether the order code of the spare part set appears in the list of the spare
parts displayed:
 YES: The spare part set may be used for the measuring device.
NO: The spare part set may not be used for the measuring device.
If you have any questions please contact your Endress+Hauser Service
organization.
5. On the Spare parts tab click the PDF symbol in the MH column.
 The Installation Instructions attached to the listed spare part are opened as a PDF
file and can also be saved as a PDF file.
6. Click one of the drawings shown on the Spare part drawings tab.
 The corresponding exploded drawing is opened as a PDF file and can also be
saved as a PDF file.
10.3

Repairing Ex-certified devices

If repairing Ex-certified devices remember that:
• Ex-certified devices may only be repaired by experienced and skilled staff or by
Endress+Hauser Service
• observe all applicable standards, certificates, national Ex-area regulations and all Safety
Instructions (XA)
• use only genuine spare parts from Endress+Hauser
• note the device designation on the nameplate to order the spare parts
• replace the component by the same type
• carry out the replacing in accordance with the instructions
Repair
75