Liquicap M FTI51
Endress+Hauser
4 →
7
3
-
+
5
> 2 s
A0042493
‣
Press the key > 2 s to set
the MIN fail-save mode.
7 →
7
3
-
+
5
A0042498
‣
The MIN-SIL mode is set.
To unlock the SIL mode and set the MAX fail-save mode (only with electronic insert
FEI55):
1 →
7
3
-
+
5
A0042483
4 →
7
3
-
+
5
> 4 s
A0042499
‣
Press the and keys > 4 s.
5 →
7
3
-
+
5
A0043657
‣
The MIN fail-save mode is set.
8 →
7
1
7
3
-
+
5
A0042632
‣
Set the function switch to
position 1.
2 →
1
7
7
3
-
+
5
A0042484
‣
Set the function switch to
position 7.
5 →
7
3
-
+
5
A0043657
‣
The SIL mode is unlocked.
Commissioning
6 →
7
3
-
+
5
> 4 s
A0042497
‣
Press the and keys > 4 s.
9
7
3
-
+
5
A0043660
3 →
7
3
-
+
5
A0042496
6
7
1
7
3
-
+
5
A0042649
‣
Set the function switch to
position 1.
55