4.2.2
AC Voltage Measurement
Measurement
Range
Specified
Measurement
(2)
Range
Accuracy
Resolution
Input Impedance
Note (1)
The display indicates "OL" above 1000V (1400V in PEAK mode).
Above 1000V (RMS), a repetitive beep indicates that the voltage
being measured is greater than the safety voltage for which the
device is guaranteed. The display indicates "OL".
- Bandwidth in AC = 3 kHz
Note (2)
Any value between zero and the min. threshold of the measurement
range (0.15V) is forced to show "----" on the display.
Specific Specifications in MAX/MIN mode
(from 10Hz to 1kHz, and from 0.30V in AC):
•
Accuracy: add 1% R to the values of the table above.
•
Capture of the extreme: approximately 100ms.
4.2.3
AC+DC Voltage Measurement
Measurement Range (2)
Specified Measurement
Range
Accuracy
Resolution
Input impedance
Note (1)
The display indicates "OL" above 1000V (1400V in PEAK mode).
Above 1000V (DC or RMS), a repetitive beep indicates that the
voltage being measured is greater than the safety voltage for which
the device is guaranteed.
- Bandwidth in AC = 3 kHz
Note (2) Any value between zero and the min. threshold of the measurement
range (0.15V) is forced to "----" on the display.
0.15 to
59.99V
0 to 100% of the measurement range
0.15 to 5.99V
± (1% R +10cts)
6.00 to 59.99V
± (1% R +3cts)
0.01V
0.15 to
59.99V
0 to 100% of the measurement range
0.15 to 5.99V
± (1% R +10cts)
6 to 59.99V
± (1% R +3cts)
0.01V
60.0 to
599.9V
± (1% R +3cts)
0.1V
10MΩ
60.0 to
599.9V
0.1V
10MΩ
36
600V to 1000V RMS
1400V peak
1V
600V to1000V RMS
1400V peak
± (1% R +3cts)
1V
(1)
(1)