2.2.2 Exhaust and drain piping

(1) Exhaust treatment
outdoor place.
(2) Drain treatment
Discharge the exhaust gas to an appropriate outdoor point.
Implement heat insulation treatment, if there is a fear of occurrence of freeze-up in the cold sea-
son.
2.2.3 Preparation of standard gas
Zero gas
Span gas
for other than
O
2
Span gas
for O
2
Case of analyzer
without O
sensor
2
N
gas or dry air
2
* A standard gas for zero
gas is not required for
atmosphere calibration
specification.
Gas of concentration
higher than 90% of full
scale
Case with
magnetic O
sensor
2
N
gas
2
Gas of concentration
higher than 90% of full
scale
Gas of concentration
higher than 90% of full
scale or atmosphere
(21%)
24
Case with
zirconia O
sensor
2
Dry air or gas of
concentration
higher than 80% of
O
range
2
* A standard gas for zero
gas is not required for
atmosphere calibration
specification.
Gas of concentration
higher than 90% of full
scale
1 to 2% O
gas
2
INZ-TN2ZSJ-E