Fuji Electric ZSJ Instruction Manual page 104

Infrared gas analyzar
Table of Contents

Advertisement

7. SPECIFICATION
Measuring system:
NOx, SO
, CO and CO
2
O
; Zirconia type, magnetic type
2
Measurable component and min./max. measurement
range:
NO
; 0 to 50 ppm/0 to 5000 ppm
X
SO
; 0 to 50 ppm/0 to 5000 ppm
2
CO ; 0 to 50 ppm/0 to 5000 ppm
CO
; 0 to 10 %/0 to 20%
2
O
; 0 to 10 %/0 to 25%
2
Number of measurement ranges:
2 Maximum range ratio: 1:10 (Refer to Code Symbols.)
Warm-up time:
Within 4 hours after power-on
Analog output signals:
Simultaneous output of signals of 4 to 20 mA DC each
(non-isolated or isolated depending on customer's code
selection)
• Five instantaneous value outputs (NO
and O
)
2
• Three instantaneous values (NO
correction when provided with O
• Three average values (NO
tion when provided with O
Contact output:
(1) Each SPST contact (contact capacity
250 V AC, 2 A or 30 V DC, 3 A) for:
• Range identification of each component (Close/1st
range), analyzing block error, calibration error, auto
calibration status, maintenance status,and CO peak
count alarm
(2) Each SPDT contact (contact capacity 250V AC, 1 A
or 30 V DC, 1 A) for:
• Concentration alarm for each component's instantaneous
Contact input:
Non-voltage contact (1.5 sec or longer)
• Auto calibration start, average value resetting
Non-voltage contact (Status holded)
• Range changeover (1st range when contact closes),
output hold, remote aspirator OFF (OFF when contact
closes)
Indication:
LCD with back light for indicating:
• Instantaneous values (NO
• O
corrected instantaneous values
2
(NO
, SO
, CO) after O
X
2
O
sensor
2
• O
corrected average values (NO
2
correction when provided with O
• O
average value when provided with O
2
• Peak count value (when provided with CO, O
• Parameter assignment
Fluorescent lamp in cubicle:
Standard equipment
Recorder (option):
Paperless recorder (Fuji Electric's type PHR) mounted
INZ-TN2ZSJ-E
; Ndir type infrared
2
, SO
, CO, CO
X
2
, SO
, CO) after O
X
2
sensor
2
, SO
, CO) after O
correc-
X
2
2
sensor
2
, SO
, CO, CO
and O
X
2
2
correction when provided with
2
, SO
, CO) after O
X
2
sensor
2
sensor
2
analyzer)
2
Gas extractor:
• Flange : JIS 5K 65AFF
• Mass: Approx. 9 kg (excluding gas sampling pipe)
• Power supply voltage: 100 V AC, 50/60 Hz
• Power consumption: Approx 100 VA
• Sampling pipe: Refer to Code Symbols for materials
and length of the pipe.
nium (length 600, 800, 1000 mm),or SiC (length 700,
900 mm)
Sample inlet tube:
cases.
2
(2) SO
of 0 to 50 ppm or 0 to100 ppm
2
(3) Tube length 10 m or longer in SO
2
(Power supply voltage: 100 V AC, 50/60 Hz, power con-
sumption: 36 VA/m)
Rated operating conditions:
tomer's code selection)
• Ambient humidity: 90% RH or less
• Power supply voltage: 100, 110, 115, 200 or 230 V AC
±10%
(depending on customer's code selection)
• Frequency: 50 or 60 Hz ±0.5 Hz
• Power consumption: Max. 900 VA
(excluding gas extractor and heating tube)
Storage condition:
drain pot should be drained before storage.)
• Ambient humidity; 95%RH or lower
Dry air:
Pressure; 100 kPa to 400 kPa
Dust and mist; None
External dimensions (H x W x D):
Indoor type; 1710 × 800 × 615 mm
Outdoor type; 1780 × 815 × 700 mm
Mass:
Approx. 300 kg (excluding standard gas)
)
2
Munsell 5Y7/1 semi-gloss
Cubicle structure:
Indoor or outdoor installation, of selfstanding type, single-
2
swing front door, plate thickness 2.3 mm standard (both
cubicle and door)
Other:
Six standard gas cylinders (3.4 L) accommodatable
Measurement Law type approval No.:
SAN181(NO
X
SAS182(SO
2
SAC182(CO analyzer)
SE171(Zirconia O
SF172(Magnetic O
95
measurement
2
analyzer)
analyzer)
sensor)
2
sensor)
2