Fuji Electric ZSJ Instruction Manual page 64

Infrared gas analyzar
Table of Contents

Advertisement

(5) The value is highlighted, indicating that
and then move the cursor to the right digit
(6) After the value is changed, press the
Meaning of setting
The setting is expressed as 1/1 full scale range
for both respective ranges.
When 0 to 1000 ppm is selected as the range,
and 10% FS is selected as hold
setting, the output equivalent to 100 ppm is
held irrespective of the measure-ment value at
that time.
(7)
eter setting screen.
Description of setting
Response time
The response time of the electrical system can
of the setting time.
INZ-TN2ZSJ-E
or the
ESC
Parameter
Hold
Ch1
Ch2
ENT
Ch3
Ch4
Ch5
Parameter Setting screen
-
2
Parameter
Response Time
Ch1
Ch2
Ch3
Ch4
Ch5
55
ENT
Set Hold value
0 to 100%FS
NOx
010
%FS
SO
020
%FS
2
CO
015
%FS
2
CO
012
%FS
O
022
%FS
2
ENT
End of Hold Setting
ESC
Select Ch No.
NOx
10
Sec.
SO
20
Sec.
2
CO
15
Sec.
2
CO
12
Sec.
O
22
Sec.
2