Operation by setting
Auto calibration is performed under the following rules.
1. Zero calibration is performed at the same time, for the Ch (component) in which "enable"
is selected at the time of auto calibration and auto zero calibration.
2. Span calibration is performed in the order from smallest Ch No., for the Ch (component)
for which "enable" is selected at the time of auto calibration.
Note
ZERO calibration on a
INZ-TN2ZSJ-E
key to exit automatic calibration component/range setting, and the previous
37