Fuji Electric ZSJ Instruction Manual page 72

Infrared gas analyzar
Table of Contents

Advertisement

(2) When press the
memorized.
of the sample gas are set automatically
is carried out.
Do not change the setting time
unless it is required.
Explanation
Set up switching time and measuring time of the reference gas and the switching time and measuring
time of the sample gas.
Switching time of reference gas : 1 to 30s (Initial value 5 sec)
Flow time of reference gas
Note) Switching time of the reference gas, measuring time of the sample gas, flow time of the
reference gas and flow time of the sample gas are updated automatically based on
following relational expression.
Switching time of sample gas (t1) = Switching time of reference gas (t3)
Flow time of sample gas (Ts) = Flow time of reference gas (Tr)
Flow time of reference gas (Tr) =
Flow time of sample gas (Ts) =
1 cycle = flow time of reference gas (Tr) + flow time of sample gas (Ts)
INZ-TN2ZSJ-E
ENT
: 1 to 60s (Initial value 5 sec)
Switching time of reference gas (t3) + measuring time of reference gas (t4)
Switching time of sample gas (t1) + measuring time of sample gas (t2)
Maintenance Mode
Sample Switch Setting
Gas Flow Time
Reference Gas
Reference Gas
Reference Gas
Sample Gas
Sample Gas
Sample Gas
Maintenance Mode
Sample Switch Setting
Gas Flow Time
Reference Gas
Reference Gas
Reference Gas
Sample Gas
Sample Gas
Sample Gas
63
Set gas replacement time.
Setting range is from 1 to
30 seconds.
Switching Time
05
s
Measuring Time
05
s
Flow Time
10
s
Switching Time
05
s
Measuring Time
05
s
Flow Time
10
s
ENT
Select an item
Switching Time
05
s
Measuring Time
05
s
Flow Time
10
s
Switching Time
05
s
Measuring Time
05
s
Flow Time
10
s