Fuji Electric ZSJ Instruction Manual page 62

Infrared gas analyzar
Table of Contents

Advertisement

Setting Range
• Hold setting
• Response time
*The cycle can be set via Maintenance Mode -> Sample Switch Setting.
One cycle (s) is the sum of the reference gas flow time (s) and sample gas flow time (s).
• Average period
• Backlight Timer
• Maintenance mode : 0000 to 9999
Output Hold
Calibration
Output hold
Auto calibration start
Calibration
Output hold
c.
Remote hold
Remote hold Input
Output hold
INZ-TN2ZSJ-E
: 0 to 100% FS
: Infrared measured component = 1 through 9 cycles (Initial value: 1 cycle)
O
meter = 1 through 60s (Initial value: 15s)
2
: 1 to 59 min or 1 to 4 hours
1 to 59 minutes when the unit is set to minute and 1 to 4 hours when it
is set to hour.
: 1 to 60 min (Initial value: 5 min)
ZERO
or
Press the
SPAN
To execute calibration, press the
(Initial value: 0000)
(Initial value: 1 hour)
(Initial value: 0000)
ENT
key to flow gas.
ENT
Hold extending
time
Time set to gas flow time (See Section 3.4 Auto Calibration.)
Auto calibration end
Hold for setting period which is set at gas flow times.
53
key.
Hold extending time.
With input