3.1.2 Manual range switch
then press the
(2) Move the highlight of the cursor to range
selection, and then select a desired range
range.)
surement is carried out in the selected
range.
Note) If "RR" or "AR" is selected as range
The ranges for O
O
correction average value, and
2
O
average value are automatically
2
switched according to the instanta-
neous value range switch settings.
(Same as for "RR" or "AR".)
To close the setting
ESC
Press the
switch mode or range switch operation or stop
the operation in the middle. The setting
operation is made invalid and the previous
screen appears.
Range identification contact operation
The range identification contact output corresponding to each Ch (component) is closed
when Range 1 is active, and open when Range 2 is active, no matter.
If the measurement value is held by remote contact input or during calibration routine and
range switch conditions are met, the contact will change position only after the hold condition
is removed.
ENT
or the
ENT
correction value,
2
key to end the setting of range
Switch Range
Ch1
MR
NOx
Ch2
AR
SO
2
Ch3
RR
CO
2
Ch4
MR
CO
Ch5
MR
O
2
Switch Range
Ch1
MR
NOx
Ch2
AR
-
SO
2
Ch3
RR
CO
2
Ch4
MR
CO
Ch5
MR
O
2
30
Select Ch No.
Switch ranges
with UP/DOWN and ENT
Back with ESC
Range1
0–50.00
Range2
0–500.0
Range1
0–50.00
Range2
0–500.0
Range1
0–10.00
Range2
0–20.00
Range1
0–50.00
Range2
0–500.0
Range1
0–10.00
Range2
0–25.00
ENT
Select Ch No.
with UP/DOWN and ENT
Back with ESC
Range1
0–50.00
Range2
0–500.0
Range1
0–50.00
Range2
0–500.0
Range1
0–10.00
Range2
0–20.00
Range1
0–50.00
Range2
0–500.0
Range1
0–10.00
Range2
0–25.00
ENT
(
)
End of Range Switch
INZ-TN2ZSJ-E
ppm
ppm
ppm
ppm
vol%
vol%
ppm
ppm
vol%
vol%
ppm
ppm
ppm
ppm
vol%
vol%
ppm
ppm
vol%
vol%