Fuji Electric ZSJ Instruction Manual page 54

Infrared gas analyzar
Table of Contents

Advertisement

cases.
Example
Start Time
Cycle
Flow Time
ON/OFF
In case where auto calibration is carried out at the above setting.
Sunday
12:00
Cycle
ChI
Zero
Span
calibration
calibration
350sec
350sec
Flow time
Setting range
Cycle
: 1 to 99 hours or 1 to 40 days (initial value 7 days)
Flow time
: 60 to 900 sec
Caution
Remote start
input.
Remote start input
INZ-TN2ZSJ-E
12:00
SUN
day
1
350 sec
Zero
350 sec
Ch1 Span
350 sec
Ch2 Span
350 sec
Ch3 Span
300 sec
Ch4 Span
300 sec
Ch5 Span
300 sec
EX. time
ON
Monday
12:00
Ch2
Ch3
Ch4
Span
Span
Span
calibration
calibration
calibration
350sec
350sec
(initial value 300 sec *180s only with CO meter)
Tuesday
12:00
Ch5
Replace-
Span
ment time
calibration
300sec
300sec
300sec
With input (hold at least 1.5 sec.)
Without input
45
: Auto calibration