3.3.2 Hysteresis setting
To prevent chattering of an alarm output near the alarm setting values, adjust the value of hyster-
esis.
(1) In the "Alarm Setting" screen that ap-
pressing the
ENT
right.
(2) Then, enter hysteresis values.
For the value entry, 1-digit value is
or the
moves the digit. After setting, press the
ENT
To close "Hysteresis Setting"
To close the "Hysteresis Setting" or cancel
the mode midway, press the
A previous screen will return.
Setting range
0 to 20% of full scale
[% full scale (% FS)] represents the
percentage with the width of the component
measurement range regarded as 100%.
Hysteresis (In case of upper limit alarm)
Hysteresis
width
Alarm
Contact output
or the
ESC
key.
Alarm Setting
Alarm-1
Alarm-2
Alarm-3
Alarm-4
Alarm-5
Hysteresis
End of Hysteresis Setting
Note
40
Set Hysteresis
0 to 20%FS available
0
0
%FS
ENT
(
)
Upper limit value
ON (closed)
OFF (open)