Fuji Electric ZSJ Instruction Manual page 49

Infrared gas analyzar
Table of Contents

Advertisement

3.3.2 Hysteresis setting
To prevent chattering of an alarm output near the alarm setting values, adjust the value of hyster-
esis.
(1) In the "Alarm Setting" screen that ap-
pressing the
ENT
right.
(2) Then, enter hysteresis values.
For the value entry, 1-digit value is
or the
moves the digit. After setting, press the
ENT
To close "Hysteresis Setting"
To close the "Hysteresis Setting" or cancel
the mode midway, press the
A previous screen will return.
Setting range
0 to 20% of full scale
[% full scale (% FS)] represents the
percentage with the width of the component
measurement range regarded as 100%.
Hysteresis (In case of upper limit alarm)
Hysteresis
width
Alarm
Contact output
or the
ESC
key.
Alarm Setting
Alarm-1
Alarm-2
Alarm-3
Alarm-4
Alarm-5
Hysteresis
End of Hysteresis Setting
Note
40
Set Hysteresis
0 to 20%FS available
0
0
%FS
ENT
(
)
Upper limit value
ON (closed)
OFF (open)
INZ-TN2ZSJ-E
-