values (zero and span). For value entry,
press the
or the
the
ENT
the
as O
2
and select the concentration listed
on the cylinder as required.
To close the setting
To close the calibration concentration value
setting process or cancel this mode midway,
ESC
press the
key.
A previous screen will return.
Setting range of values
NOx, SO
, CO
CO, external O
2
2,
measurement and built-in O
External Zirconia O
measurement
2
The setting cannot be performed beyond the range.
Span gas: 1 to 105% of full scale (Full scale (FS)
2
sensor
is the same as each range value.)
2
Zero gas: 5 to 25 vol% / Span gas: 0.01 to 5 vol%
32
Cursor for setting value
Cal. Settings
Set calibration value
Cal. Value
CH
RANGE
Ch1
0–50.00ppm
NOx
0–500.0ppm
Ch2
0–50.00ppm
SO
0–500.0ppm
2
Ch3
0–10.00ppm
CO
0–20.00ppm
2
Ch4
0–50.00vol%
CO
0–500.0vol%
Ch5
0–10.00vol%
O
0–25.00vol%
2
End of Calibration
Concentration Setting
ENT
(
)
ZERO
SPAN
+0000.0
50.00
+00000
500.0
+0000.0
50.00
+00000
500.0
+000.00
10.00
+000.00
20.00
+0000.0
50.00
+00000
500.0
21.00
01.00
21.00
01.00
ENT
INZ-TN2ZSJ-E