Fuji Electric ZSJ Instruction Manual page 44

Infrared gas analyzar
Table of Contents

Advertisement

3.2.3 Setting of calibration range
ing the
or the
• If "current" is selected, zero or span cali-
Press the
To close "Setting of Calibration Range"
To close "Setting of Calibration Range" or to cancel this mode midway, press the
A previous screen will return.
Example
Ch1
Range 1: 0 to 50 ppm
NOx
Range 2: 0 to 500 ppm
Ch2
Range 1: 0 to 50 ppm
Range 2: 0 to 500 ppm
SO
2
Ch1: Range 1 and Range 2 are calibrated together.
Ch2: Only currently displayed range is calibrated.
Note
To perform calibration for
"both," set the same calibra-
tion gas concentration for
both ranges.
INZ-TN2ZSJ-E
or the
ENT
Cell. Settings
Cell. Range
Ch1
-
NOx
Ch2
ENT
SO
2
Ch3
CO
2
Ch4
CO
Ch5
-
O
2
End of Calibration Range Setting
Manual Calibration screen
When setting NOx and CO to "both"
ZERO Call.
Ch1
Range1
Range2
NOx
Ch2
Range1
SO
Range2
2
Ch3
Range1
CO
Range2
2
Ch4
Range1
CO
Range2
Ch5
Range1
O
Range2
2
Two cursors will appear in both ranges (Ch1 and Ch4).
35
Set calibration range
current or both range
Range1
0–50.00
ppm
Range2
0–500.0
ppm
Range1
0–50.00
ppm
Range2
0–500.0
ppm
Range1
0–10.00
vol%
Range2
0–20.00
vol%
Range1
0–50.00
ppm
Range2
0–500.0
ppm
Range1
0–10.00
vol%
Range2
0–25.00
vol%
ENT
(
)
ESC
key.
both
current
ENT : Go on Calibration
of selected Ch
ESC : Not calibration
-0.6
0–50.00
ppm
0–500.0
ppm
0.4
0–50.00
ppm
0–500.0
ppm
0.00
0–10.00
vol%
0–20.00
vol%
-0.1
0–50.00
ppm
0–500.0
ppm
0–10.00
vol%
21.00
0–25.00
vol%
both
current
current
both
current