6.2 Troubleshooting for analyzer unit
Error display
Error No.1
Light source/motor rotation is
faulty.
Error No.2
Detector failure
Error No.3
A/D error
Error No.5
Amount of zero calibration
(indication value) is over 50%
of full scale.
Error No.6
Error No.7
Amount of span calibration
and calibrated concentration) is over
50% of full scale.
Error No.8
Measured values fluctuate too much
during zero and span calibration.
Error No.9
calibration.
Error No.10
Output cable connection is
improper.
missed.
INZ-TN2ZSJ-E
Error contents
Probable causes
Infrared light source is faulty.
Sector motor is not properly run or is stopped.
Amplifier circuit is faulty.
Detector voltage circuit is faulty.
Amplifier circuit is faulty.
A/D conversion circuit is failure.
match cylinder concentration.
occurred during auto calibration.
DIO circuit is failure.
93
.