Door Interlocks - GE Innova 2100-IQ Preinstallation Manual

Cardiovascular imaging system
Hide thumbs Also See for Innova 2100-IQ:
Table of Contents

Advertisement

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare

9 Door Interlocks

IEC 60601–2–43 requires not to install door interlocks. It is the responsibility of
the field service to check that this requirement is not in contradiction with local
regulation. In case of conflict, follow local regulation. No other measures employed
for radiation protection should cause the interruption of irradiation and any other
disturbance of a procedure in progress.
The door interlocks is wired to the C2 cabinet.
Chapter 6 Electrical Connections
Direction 5314528-2-1EN, Revision 1
175

Advertisement

Table of Contents
loading

This manual is also suitable for:

Innova 3100Innova 3100-iqInnova 4100Innova 4100-iq

Table of Contents