Section 3.0 Delivery Requirements
3.1
Door Size Requirements
Minimum door sizes also apply to hallway and elevator.
Door Height: The minimum door height accommodated is 190 cm (75in) when the Wall Stand
is tilted on the dolly.
Door Width:
• The minimum door width to accommodate the Table is: 95cm (37.5in).
• The minimum door width is calculated based on a straight-in approach requiring a 2.5 m (8.20 ft)
wide corridor. Minimum widths will change based on narrower corridors.
• The minimum door width to accommodate the Image Paste Barrier is 1.1m (44in) wide x 2.1m (82in) high.
3.2
Minimum Elevator Depth Requirements
The minimum elevator depth to accommodate is 2.45 m (96.46 in) when the Wall Stand is tilted on
the dolly.
3.3
Shipping Fixtures and Carts
The wall stand is delivered on a fixture.