Lca Error Codes On The Ena - Agilent Technologies N4373A User Manual

Lightwave component analyzer
Table of Contents

Advertisement

Troubleshooting and Maintenance

LCA Error Codes on the ENA

96
Table 5-1.
Error
Error Contents
Number
904
Unable to locate the O/E
characteristic file S21.
905
Unable to locate the O/E
characteristic file S22.
911
The starting frequency in the
specified calibration data differs
from the electrical calibration
data.
912
The ending frequency in the
specified calibration data differs
from the electrical calibration
data.
913
The measurement points in the
specified calibration data differ
from the electrical calibration
data.
999
File recall operation was
cancelled.
• Please reference the attached error code table if an error
messages is displayed.
• The last three digits in the error message are the error
code.
Error
Error Contents
Number
101
Average Number Input Error
102
Channel Number Input Error
201
Calibration Data(ErrTerm) cannot
Dounload
Agilent N4373A Lightwave Component Analyzer, First Edition
Comments / Response
Add the O/E characteristic file S21
to the RefFiles folder.
Add the O/E characteristic file S22
to the RefFiles folder.
Displayed during file selection for
O/E and O/O measurements.
Verify the selected measurement
data and current calibration data.
Displayed during file selection for
O/E and O/O measurements.
Verify the selected measurement
data and current calibration data.
Displayed during file selection for
O/E and O/O measurements.
Verify the selected measurement
data and current calibration data.
Displayed during file selection for
O/E and O/O measurements.
Displayed when [Cancel] is
selected during file selection.
Comments / Response
Enter an average value within the
acceptable range (numerical
value).
Input the channel number within
the acceptable range (numerical
value).
Calibration may not have been
performed. Perform calibration.