Specifications and Regulatory Information
116
System dynamic range Ratio of electrical powers, expressed
in dBe. The definition of system dynamic range depends on
the type of measurement:
O/O Difference between the electrical power in dBm
measured with
LCA optical in and the
electrical in)
applied to LCA optical in).
Measurement: after O/O setup calibration do a
measurement with darkened LCA optical in.
O/E Difference between the
power
in dBm and the receiver
(of LCA electrical in)
electrical in).
Condition: wavelength as specified.
Note: O/E system dynamic range is identical to E/E
system dynamic range.
E/O Difference between the electrical power in dBm
measured with the largest possible
power
at the LCA optical in, and the
floor (of LCA electrical in)
Conditions: maximum
LCA optical in as specified.
Note: The largest possible modulated optical power at
the LCA optical in is defined by the
operating input power (of LCA optical
Measurement:
1 Measure the
receiver (dB
electrical power (dBm) corresponding to the
Maximum linear operating input power (of LCA
optical
in).
2 Measure the electrical noise floor (dBm) with
darkened LCA optical in.
3 Calculate the difference between both (expressed in
dBe).
Agilent N4373A Lightwave Component Analyzer, First Edition
Modulated optical power
Electrical noise floor (of LCA
in dBm (when no optical power is
Maximum LCA electrical in
Electrical noise floor
in dBm (with terminated LCA
in dBm.
Modulated optical power
Responsivity
of the LCA optical
). Using this value, calculate the
A/W
Introduction
applied on
Modulated optical
Electrical noise
at the
Maximum linear
in).