Agilent Technologies N4373A User Manual page 115

Lightwave component analyzer
Table of Contents

Advertisement

Specifications and Regulatory Information
N O TE
N O TE
114
Measurement: Zero output is created by the following:
• for O/E: the LCA electrical in port is terminated with 50
Ohm load.
• for E/O: the LCA electrical out is terminated with 50 Ohm
load, LCA optical in is darkened.
• for O/O: the LCA optical in is darkened.
The average is calculated within frequency sections of 1/10
of the specified system bandwidth. The result must be in
limit in each section.
Modulated optical power The amplitude of a harmonic optical
power modulation. The modulated optical power is calculated
from the peak optical power P
power P
,
avg
in Watts: P
= P
mod
in dBm:
P
= 10⋅log(P
mod,dB
P
opt
P
P
Modulated optical power divided by average optical power is a value
between 0 (no modulation) and 1 (maximum modulation).
Optical noise floor (LCA optical in) The optical equivalent (see
Modulated optical
power) of the
LCA electrical in)
(with darkened LCA optical in), expressed
in dBm.
The optical equivalent depends on the
within the optical test set.
Responsivity The amplitude response R of an E/O or O/E
converter to a harmonic stimulus, expressed in units of...
Agilent N4373A Lightwave Component Analyzer, First Edition
and the average optical
max
– P
max
avg
/ 1mW).
mod
(t)
harmomic
oscillation
max
avg
Electrical noise floor (of
Responsivity
Introduction
P
mod
t
of the optical receiver