Returning to Electrical Measurements on the ENA
Agilent N4373A Lightwave Component Analyzer, First Edition
You should only perform this calibration if you have already
calibrated for E/O, O/E or O/O measurements.
1 Press the "Macro Run" button on the front panel of the
network analzer.
Select [E/E Measurement]
The following dialog box will be displayed:
2 Make sure the same measurement channel is selected as
you used for the initial electrical calibration (and the
optical measurements you have made since then).
3 If you are using an adaptor to connect a device with an
output with 75 Ω impedance to the 50 Ω input of the
ENA, please check the box here.
4 Click on the [Start] button to start the measurement
preparation.
5 When the message box is displayed, click the [OK] button
to start the measurement.
Please refer to the user guide and online help of the
network analyzer for further information on electrical
measurements.