Agilent Technologies N4373A User Manual page 98

Lightwave component analyzer
Table of Contents

Advertisement

Agilent N4373A Lightwave Component Analyzer, First Edition
Error
Error Contents
Number
203
O/E characteristic File S21 Read
Error
204
O/E characteristic File S22 Read
Error
205
Display Trace Count Over
206
O/E characteristic File
is different S21 and S22
207
Calibration Data(ErrTerm) File
Read Error
401
ENA(E5071B) not Response
901
Average File (Average.txt) not
Found
902
Average File(Average.txt) Read
Error
903
Average File(Average.txt) Write
Error
904
O/E characteristic File S21 not
Found
905
O/E characteristic File S22 not
Found
911
Different Start Frequency
(Between Select File and ErrTerm)
912
Different Stop Frequency
(Between Select File and ErrTerm)
Troubleshooting and Maintenance
Comments / Response
The specified O/E characteristic
file may not exist. Verify the file.
The specified O/E characteristic
file may not exist. Verify the file.
The displayable trace count has
exceeded its limit. Delete the
trace count.
The S21 and S22 file attributes
(sweep type, wavelength, NOP)
are different. Verify the contents
of the files.
Output during an E/E Measure.
Other measurements may not
have been performed in advance.
The GPIB setting for the ENA may
not be configured correctly. Verify
the configuration.
Add the average file to the
Conditions folder.
The average file may not be
properly specified.
The average file may be write-
protected. Change the write
permissions as necessary.
Add the O/E characteristic file S21
to the RefFiles folder.
Add the O/E characteristic file S22
to the RefFiles folder.
Displayed during file selection for
O/E and O/O measurements.
Verify the selected measurement
data and current calibration data.
Displayed during file selection for
O/E and O/O measurements.
Verify the selected measurement
data and current calibration data.
97

Advertisement

Table of Contents
loading

Table of Contents