Agilent Technologies N4373A User Manual page 127

Lightwave component analyzer
Table of Contents

Advertisement

Specifications and Regulatory Information
126
Number of averages
Smoothing
S Parameter to measure
Measurement mode
LCA file format
Optical power meter averaging time
avg
T
Option #300
Frequency sweep range
RF output power
Option #301
Frequency sweep range
RF output power
Option #302
Frequency sweep range
RF output power
Option #303
Frequency sweep range
RF output power
2 Perform a 2- port electrical calibration of the network
analyzer, including the LCA RF cables, according to the
User's Guide of that instrument. For convenience, use of
an ECal module is recommended. Ports 1 and 2 are used
for these procedures. After this calibration, the RF cables
should not be removed from the network analyzer ports.
The free end of the RF cable from Port 1 will be referred
to as LCA RF Output. The free end of the RF cable from
Port 2 will be referred to as LCA RF Input.
Agilent N4373A Lightwave Component Analyzer, First Edition
Introduction
1
None
S21, S22
Single ended
.S2P "Log Mag/Angle
(dB/degrees)"
1 s
Whole NWA frequency range (300
kHz...3 GHz)
Max. (leveled) NWA output power
over whole frequency range (+10
dBm)
Whole NWA frequency range
(300 kHz ... 8.5GHz)
Max. (leveled) NWA output power
over whole frequency range (+5
dBm)
Whole NWA frequency range
(12.5 MHz ... 20 GHz)
Max. (leveled) NWA output power
over whole frequency range (+3
dBm)
Whole NWA frequency range
(12.5 MHz .. 20 GHz)
Max. (leveled) NWA output power
over whole frequency range (-3
dBm)

Advertisement