Performance Verification on the PNA

26
To check the correct operation of the Lightwave Component Analyzer:
1 Turn on the network analyzer and the optical test set.
2 Allow the Lightwave Component Analyzer to warm up,
that is until the LEDs for "LD Temp Ready" and "PD
Temp Ready" are lighting.
3 Turn on the laser by turning the key to the upright
position.
The "On" LED lights to indicate the laser is on.
4 If your Lightwave Component Analyzer has more than one
wavelength, select the wavelength at which you want to
measure by pressing the button under the Wavelength
LEDs.
5 Start the network analyzer application.
Consult the documentation supplied with the network
analyzer if you need help with this step.
6 Perform an electrical calibration, as described in
"Calibrating the network analyzer before measurements"
on page 56.
7 Switch on the laser of the optical test set, with the key
switch.
8 Select [System] > [Macro] > [O/O Measurement].
Agilent N4373A Lightwave Component Analyzer, First Edition