Agilent N4373A Lightwave Component Analyzer, First Edition
To prevent electrical shock, disconnect the instrument from mains
before cleaning. Use a dry cloth or one slightly dampened with
water to clean the external case parts. Do not attempt to clean
internally.
Using controls or adjustments or performing procedures other
than those specified in the documentation supplied with your
equipment can result in hazardous radiation exposure.
This product complies with overvoltage Category II and Pollution
Degree 2.
VENTILATION REQUIREMENTS: When installing the product in a
cabinet, the convection into and out of the product must not be
restricted. The ambient temperature (outside the cabinet) must be
less than the maximum operating temperature of the product by 4 °C
for every 100 watts dissipated in the cabinet. If the total power
dissipated in the cabinet is greater than 800 watts, then forced
convection must be used.
Install the instrument so that the Power switch is readily identifiable
and is easily reached by the operator. This is the instrument
disconnecting device. It disconnects the mains circuit from the
mains supply before other parts of the instrument. Alternatively, an
externally installed switch or circuit breaker (which is readily
identifiable and is easily reached by the operator) may be used as a
disconnecting device.
Always use the three-prong AC power cord supplied with this
instrument. Failure to ensure adequate earth grounding by not using
this cord may cause instrument damage.
This instrument has autoranging line voltage input. Be sure the
supply voltage is within the specified range.