Making Measurements on the PNA
58
Electrical calibration for balanced measurements
If you are measuring multiport devices, calibrate the network
analyzer for balanced measurements before you start using
the Lightwave Component Analyzer, or return to E/E
measurement mode, as described in "Returning to Electrical
Measurements on the PNA" on page 75, to do the calibration.
1 Perform a 3 port electrical calibration.
a Start the Calibration wizard
b Select "SmartCal", and click [Next >].
c Select "3 Port Cal" and click [Next >].
d Select the appropriate Cal Kit for each port, and click
[Next >]..
e Follow the instructions on the screen.
Agilent N4373A Lightwave Component Analyzer, First Edition