Measuring Opto-Electrical Devices on the PNA
Calibrating the O/E measurement
Agilent N4373A Lightwave Component Analyzer, First Edition
Always perform an electrical calibration of the network
analyzer or load an existing calibration file before you make
a measurement. If you make a measurement without this
calibration, you will get an error message.
For standard measurements, see "Electrical calibration for
single- ended measurements" on page 57.
For balanced measurements, see "Electrical calibration for
balanced measurements" on page 58.
Before you make your first opto- electrical measurement, you
must calibrate the O/E measurement.
If you have already calibrated the O/E measurement, you can
connect your device under test and make the measurement
using the controls on the network analyzer.
Please refer to the user guide and online help for the
network analyzer for further information.
1 Select [System] > [Macro] > [O/E Measurement]
The following dialog box will be displayed:
Making Measurements on the PNA
65