Op Panel Trackball Assy - GE LOGIQ 5 Service Manual

Hide thumbs Also See for LOGIQ 5:
Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2300000, R
IRECTION
EVISION
8-4-9

OP Panel Trackball Assy

Purpose: This is a description on how to remove and replace OP Panel Trackball Assy.
8-4-9-1
Tools
Common pillips screwdrivers.
8-4-9-2
Needed Manpower
1person, 5 minutes + travel
8-4-9-3
Preparations
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 3.
Keyboard Assy should be removed before proceed. For more information, refer to KeyBoard Assy
on section 8-4-1 on page 15.
8-4-9-4
Removal Procedure
1.) Disconnect two connectors (1-2) from the Trackball Assy. Refer to Figure 8-37.
2.) Unscrew three screws (3-5) to remove the Trackball Assy. Refer to Figure 8-38.
2
2
Figure 8-37 Disconnect two connectors
4
5
Figure 8-38 Unscrew 3 screws
Chapter 8 Replacement Procedures
LOGIQ™5 S
ERVICE
1
3
M
ANUAL
8-31

Advertisement

Table of Contents
loading

Table of Contents